Nuevo ELISA Kit for Heat Shock Protein 70 (HSP70) Ver mas grande

ELISA Kit for Heat Shock Protein 70 (HSP70)

CC-SEA873Po

Producto nuevo

48T, 96T, 96T×5, 96T×10, 96T×100

Más detalles

Por favor regístrese para ver el precio

Comprando este producto generará 7 Biopuntos. Su cesta contiene un total 7 Biopuntos puede ser convertido en un Biobonos Descuento 28.00EUR.


Añadir a Mis Favoritos

Hoja técnica

Size 48T, <i>96T</i>, 96T×5, 96T×10, 96T×100
Applications Enzyme-linked immunosorbent assay for Antigen Detection.
Detection Range 3.12-200ng/mL
Sensitivity The minimum detectable dose of this kit is typically less than 1.16ng/mL
Research Area Tumor immunity
Organism species Sus scrofa
Alternative Names DnaK
Item Name Heat Shock Protein 70
Assay Length 3h
Method Double-antibody Sandwich
Sample Type serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Test Principle The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Heat Shock Protein 70 (HSP70). Standards or samples are then added to the appropriate microti
Assay procedure summary 1. Prepare all reagents, samples and standards
UniProt ID -

Más información

Especificidad:

This assay has high sensitivity and excellent specificity for detection of Heat Shock Protein 70 (HSP70).
No significant cross-reactivity or interference between Heat Shock Protein 70 (HSP70) and analogues was observed.

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Heat Shock Protein 70 (HSP70) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Heat Shock Protein 70 (HSP70) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Estabilidad:

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Click here for manual in PDF

Click here for OD image